About
The FAA 200 combines high throughput, accuracy, and intelligent automation in a space-saving design. With random access testing, a user-friendly software interface, and smart reagent management, it ensures streamlined workflows, reduced manual intervention, and faster turnaround times for busy clinical laboratories.This analyzer also supports LIS connectivity and advanced quality control features to ensure traceability and compliance, providing dependable support for modern diagnostic environments.
Specification
Key Features:
– Test Speed: Up to 200 tests/hour (mono or double reagent)
– Sample Tray: 40 positions
– Reagent Tray: 80 positions, refrigerated (24-hour cooling)
– Reaction Cuvettes: 44 reusable cuvettes with automatic cleaning and drying
– Sample Volume: 2–30 µL (0.1 µL steps)
– Reagent Volume: 20–300 µL (1 µL steps)
– Precision: CV% ≤3% (ALT), ≤2.5% (Urea), ≤2% (TP)
– Water Consumption: <8 L/hour
Technical Advantages:
– Methods supported: Colorimetry, Turbidimetry, 1-point and 2-point end
– Calibration: Linear, Nonlinear, up to 8-point calibration
– Photometry: Wavelength range — 340nm, 405nm, 505nm, 546nm, 578nm, 630nm, 700nm
– Temperature: 37±0.1°C (Peltier-based)
– Data Interface: LIS connectivity (HL7 protocol, TCP/IP, RS232)
– Software: Windows-based, intuitive, with user-level access control
Automation Features:
– Automated reagent loading and tracking
– Paddle-type mixer to minimize carry-over
– Barcode scanning for sample ID tracking
– Air purge system to eliminate bubbles
– Built-in quality control with multi-level reports
– Alerts for reagent usage and maintenance
Direction for Use
1. Initial Setup:
– Place the analyzer on a stable, vibration-free bench.
– Connect the power supply and ensure proper grounding.
– Fill the water tank and connect drainage.
2. Startup Procedure:
– Power on and perform system self-checks.
– Check reagent and sample trays.
– Ensure cuvettes are clean and ready.
3. Loading Samples and Reagents:
– Load patient samples into the 40-position sample tray.
– Load reagents into the refrigerated reagent tray with barcode labels.
– Use only validated reagents.
4. Running Tests:
– Select tests using the Windows-based software.
– Scan barcodes for identification.
– Confirm parameters and start the automated run.
5. Post-Run Maintenance:
– Review/export results via LIS or print reports.
– Flush system with water or cleaning solution.
– Perform daily and weekly maintenance as per the manual.
6. Shutdown:
– Power down properly.
– Clean trays and wipe external surfaces.
Other Info
Dimensions & Power –
Size: 745 mm Ă— 540 mm Ă— 530 mm
Weight: ≤80 kg
Power: AC 220V/110V, 50/60Hz, 1000 VA