About
-Autozyme Urea is a reagent set for determination of Urea/Blood Urea Nitrogen based on enzymatic method using Urease
-Urea can be determined in 8 minutes
-Autozyme Urea is linear up to 350 mg%
– Autozyme Urea is a two reagent system using two step procedure
-The shelf life of Autozyme Urea is 18 months
Pack Size : 2 x 50 det. (200 ml)
Principle
Urease splits urea into ammonia and carbon dioxide. Ammonia released in this reaction reacts with hypochlorite and phenolic chromogen to produce green color. The absorbance of this green color at 578 nm. (570-620) is directly proportional to the concentration of urea in specimens.
Procedure
Reaction time …………………………. 3+ 5 min.
Wavelength …….. 578 nm. (570-620 nm.)
Zero setting with. …….. Reagent Blank
Blank absorbance limit ………… <0.200 Abs.
Sample volume….. ……. 0.01 ml (10)
Reagent volume ……………………. 1.0 ml + 1.0 ml
Standard concentration………….. 40 mg%
Linearity ……….350 mg/dl
Quality Control
To ensure adequate quality control, it is recommended that each batch should include a normal and an abnormal commercial reference control serum. It should be realized that the use of quality control material checks both instrument and reagent functions together. Factors which might affect the performance of this test include proper instrument function, temperature control, cleanliness of glassware and accuracy of pipetting.
Other Details
The reagent kit should be stored at 2-8°C and is stable till the expiry date indicated on the label. The reconstituted enzyme solution is stable for 6 months when stored at 2-8°C. DO NOT FREEZE THE ENZYME SOLUTION. The chromogen solution is stable for 6months at 2-8°C after opening.
Laxmi Sharma –
Its worth recommending u to others… Quality reagents
Aakash Kisan –
I Am using accurex company biochemistry reagents like CRP, RA, urea reagents one month. I am satisfied with Quality and service.
Rajeev Rana –
I am using your biochemistry reagents Last one year, good quality.
Raj Rajat –
Accurex from last More than 20 years it’s provide Brand Confidance in front of customer and Market leader in biochemistry kits .
Shyam Patvardhan –
We are using accurex biochemistry product’s and getting superb result from years
Gautam Rana –
We are using Accurex Biochemistry reagents, and we are really very happy with the quality of products.