ACCUDx CQ Cardiac Troponin I Kit is used for in-vitro quantitative determination of Cardiac Troponin I (cTni) in serum, plasma or whole blood. This test is used as an aid in the diagnosis of myocardial injury such as Acute Myocardial Infarction (AMI), Unstable Angina, Acute Myocarditis and Acute Coronary Syndrome (ACS).

Buy Trop I Test Kit (Pack of 10 test) for Trop I blood Testing. TnI levels are considered elevated if they’re above 40 ng/L in a high-sensitivity Troponin I test. Troponin is a molecular complex that is bound to the thin filament (actin) of striated muscle fibers.

 

4 in stock

About

ACCUDx CQ Cardiac Troponin I Kit (Immunofluorescence Assay) is intended for in vitro quantitative determination of Cardiac Troponin I (cTni) in serum, plasma or whole blood. This test is used as an aid in the diagnosis of myocardial injury such as Acute Myocardial Infarction (AMI), Unstable Angina, Acute Myocarditis and Acute Coronary Syndrome (ACS).

Troponin, a molecular complex that is bound to the thin filament (actin) of striated muscle fibers, acts with intracellular calcium to control the interaction of the thin filament with the thick filament (myosin), thus regulating muscle contraction. Troponin consists of three regulatory proteins: T, which connects the troponin complex and tropomyosin (another cardiac muscle regulatory protein); I, which prevents muscle contraction in the absence of calcium; C, which binds calcium. Cardiac Troponin I (MW 22.5 kDa) and the two skeletal muscle isoforms of Troponin I have considerable amino acid sequence homology, but cTni contains an additional N-terminal sequence and is highly specific for myocardium.
Clinical studies have demonstrated the release of cTnl into the blood stream within hours following acute myocardial infarction (AM) or ischemic damage. Elevated levels of cTnl are detectable in blood within 4 to 6 hours after the onset of chest pain, reaching peak concentrations in approximately 8 to 28 hours, and remain elevated for 3 to 10 days following AM. Due to the high myocardial specificity and the long duration of elevation, Tni has become an important marker in the diagnosis and evaluation of patients
suspected of having an AMI.
The current quidelines cf The Joint European Society of Cardiclogy/Amarican College of Cardiology Committee support the use DicTni as a preferred marker of myocardial injury. Several major studies have shown that cTni is also a predictor of cardiac risk in patients with unstable angina. The American College of Cardiology and the American Heart Association’s current guidelines recommend using troponin results when making treatment decisions regarding unstable angina and non-ST segment
elevation MI (NSTEMI).

TEST RESULTS:
ACCUDx CQ Analyzer can scan the test card automatically and display the result on the screen. For additional information, please refer to the user manual of ACCUDx CQ Analyzer”

Specification

The test uses an anti-human cTnl monoclonal antibody conjugated with fluorescence latex and another anti-human cTnl monoclonal antibody coated on the test line. After the sample has been applied to the test strip, the fluorescence latex-labelled anti-human cTni monoclonal antibody binds with the cTnl in sample and forms a marked antigen-anlibody complex. This complex moves to the test delection zone by capillary action. Then marked antigen-antibody complex is captured on the test line by the anti-human cTnl monoclonal antibody. The fluorescence intensity of the test line increases in proportion to the amount of cTnl in sample.
Then, the test card is inserted into ACCUDx CQ Immunofluorescence Quantitative Analyzer. The concentration of cTnl in sample will be measured and displayed on the screen. The value will be stored in ACCUDx CQ Analyzer. The result can be transmitted to the laboratory or hospital information system.”

Safety Information

1. For in vitro diagnostic use only.
2. For professional use only.
3. Do not use the kit beyond the expiration date.
4. Do not use the test card if the foil pouch or the cartridge is damaged.
5. Do not open pouches or the cartridge until you are ready to perform test.
6. Do not reuse the test card.
7. Do not reuse the pipet tips.
8. Mandie ail specimens as potentially infectious. Proper handing and disposal methods should be followed in accordance with local regulations and good safety practices.
9. Carefully read and follow the user manual to ensure test performance.”

Direction for Use

1. Collect specimens according to safe phlebotomy procedures.
2. Test card, sample and reagent should be brought to room temperature before testing
3. Confirm that SD card lot No. matches with the test kit lot No… Perform ‘SD Card Calib”” calibration whenever lot number is changed and when necessary.
4. On the main menu of ACCUDx CQ Analyzer, press “”ENT* button to enter testing interface.
5. Remove the test card from the sealed pouch just before use. Label the test card with patient or specimen identification.
on Standardization of Markers of Cardiac Damage. Clin Chem Lab Med,
1998, 36:887-893.
2. Antman EM, Anbe DT, Armstrong PW, et al. ACC/AHA guidelines for the management of patients with ST-elevation myocardial infarction: a report of the American College of Cardiology/American Heart Association Task
Force on Pratice Guidelines (Committee to Revise the 1999 Guidelines
for the Manage 2004).
3. EN ISO 18113-1:2009 In vitro diagnostic medical devices – Information supplied by the manufacturer (labelling) – Part 1: Terms, definitions and
general requirements.
4. EN ISO 18113-2:2009 In vitro diagnostic medical devices – Information supplied by the manufacturer (labelling) – Part 2: In vitro diagnostic
reagents for professional use (ISO 18113-2:2009).

Other Info

Store the test card at 4-30°C.
Use the lest card within 1 hour once the foil pouch is opened.
Store the whole blood buffer at 0-30°C. For better results, store the whole blood buffer at 2-8°C.

5 reviews for AccuDx CQ Trop I (10 Test)

  1. Dhanuk dhari Pakre

    Accurex is a quality based reagents manufacturing company .We are satisfied with your products. We try to more and more use of your products.

  2. Vinod Praksh

    I am using your Troponin I. hba1c., and am very satisfied..

  3. Laxmikant Lad

    Products quality very Good

  4. Kumar Rane

    Results are quite satisfactory.

  5. Sanjana Ratik

    excellent results

Add a review

Your email address will not be published. Required fields are marked *

Fill Captcha *Time limit exceeded. Please complete the captcha once again.

×

Hello!

Click below to chat with us on WhatsApp

×